Composition and Morphology of wear debris in Failed Uncemented Total Hip Replacement

Journal of Bone & Joint Surgery

Shanbhag A.S., Jacobs J.J., Glant T.T., Gilbert J.L., Black J., Galante J.O.

Abstract:

Interfacial membranes collected at revision from 11 failed uncemented Ti-allo total hip replacements were examined. Particles in the membranes were characterized by electron microscopy, microchemical spectroscopy and particle size analysis. Most were polyethylene and had a mean size of 0.53 µm ± 0.3. They were similar to the particles seen in the base resin used in the manufacture of the acetabular implants. Relatively few titanium particles were seen. Fragments of bone, stainless steel and silicate were found in small amounts. Most of the polyethylene particles were too small to be seen by light microscopy. Electron microscopy and spectroscopic techniques are required to provide an accurate description of this debris.