Improvement of Accuracy in High Capacity Load Cell: Application to Robotic Testing of Musculoskeletal Joints.

Annals of Biomechanical Engineering

Lars G. Gilbertson, Todd C. Doehring, Glen A. Livesay, Theodore W. Rudy, James D. Kang, Savio L-Y. Woo

Abstract:

This study investigated a previously unaccounted for source of error in a high-capacity, six degree-of-freedom load cell used in multi-degree-of-freedom robotic testing of musculoskeletal joints, an application requiring a load cell with high accuracy in addition to high load capacity. A method of calibration is presented for reducing the error caused by changes in universal force-moment sensor (UFS) orientation within a gravitational field. Uncorrected, this error can exceed a magnitude of 1% of the full-scale load capacity-the manufacturer-stated accuracy of the UFS. Implementation of the calibration protocol reduced this error by approximately 75% for a variety of loading conditions. This improvement in load cell accuracy (while maintaining full load capacity) should improve both the measurement and control of specimen kinetics by robotic/UFS and other biomechanical testing systems.